CyberKnife IRIS Test

Description

This test measures the size of fixed or variable (IRIS) collimation. It compares the current results to the results in commissioning.

Once the image has been imported, the analysis proceeds automatically.

The plots will show the radial and transverse profiles through the centroid of each image. FWHM measurements and difference measurements between the current and comissioning results.

Requirements

Software products
RITG135
Hardware
GafChromic Films

Additional information

Reference TG135

Example Report