Varian RapidArc®

Description

The purpose of the Varian RapidArc® MLC QA Tests is to provide six automated routines to support commissioning and ongoing QA on Varian LINACs equipped with either a Millennium 120MLC or a Millennium HD120 MLC and RapidArc®.

Test 0.1 – dMLC dosimetry
Test 0.2 – Picket Fence versus Gantry angle
Test 1.1 – Picket Fence test during RapidArc®
Test 1.2 – Picket Fence test during RapidArc® with intentional errors
Test 2 – Dose Rate and Gantry Speed during the delivery of RapidArc®
Test 3 – Leaf speed during RapidArc® delivery

RapidArc® is registered trademark of Varian Medical Systems, Inc.

Varian RapidArc® test 0.2 EPID image
Varian RapidArc® test 0.2 EPID image
Varian RapidArc® test 1.1 EPID image
Varian RapidArc® test 1.1 EPID image
Varian RapidArc® test 1.2 EPID image
Varian RapidArc® test 1.2 EPID image
Varian RapidArc® test 2.0 EPID image
Varian RapidArc® test 2.0 EPID image
Varian RapidArc® test 3.0 EPID image
Varian RapidArc® test 3.0 EPID image

Requirements

Hardware
Varian dynamic MLC
Varian aS 500/1000 EPID

Additional information

Reference TG142

Example Report: Test 0.1

Example Report: Test 0.2

Example Report: Test 1.1

Example Report: Test 1.2

Example Report: Test 2.0