Linac Conebeam CT Image QA

Description

Using RIT software and Catphan phantom will help test the following parameters of Elekta XVI and Varian OBI panels:

  • HU of CT materials
  • CT Number linearity
  • Slice Width
  • Low Contrast Sphere Detection
  • CT Uniformity
  • Low Contrast Object Detection
  • Contrast to Noise Ratio
  • Spatial Resolution and MTF
  • Horizontal and Vertical Scaling
  • Supports:

  • All Varian OBI devices, Elekta XVI
  • All TG-142 recommended kV Imaging tests
  • In addition, it provides automated tracking, trending and reporting.

    Catphan 404 analyzed image
    Catphan 404 analyzed image
    Catphan 486 analyzed image
    Catphan 486 analyzed image
    Catphan 515 analyzed image
    Catphan 515 analyzed image
    Catphan 528 analyzed image
    Catphan 528 analyzed image
    Elekta XVI 486 analyzed image
    Elekta XVI 486 analyzed image
    Elekta XVI LCV analyzed image
    Elekta XVI LCV analyzed image
    Elekta XVI resolution analyzed image
    Elekta XVI resolution analyzed image

    Requrements

    Hardware
    Varian OBI
    Elekta XVI
    Catphan 500 / 503 / 504 / 600

    Additional information

    Reference TG142

    Example Report: Catphan 404 Module

    Example Report: Catphan 486 Module

    Example Report: Catphan 515 Module

    Example Report: Catphan 528 Module

    Example Report: Catphan 732 Materials Module